ACCURACY OF ULTRA LOW DOSE AND STANDARD DOSE CBCT PROTOCOLS IN VOLUMETRIC MEASUREMENTS OF SIMULATED BONE DEFECTS VERSUS REAL MEASUREMENTS :
المؤلف
اHekmat-Allah Sabry El sawy;
هيئة الاعداد
باحث / Hekmat-Allah Sabry El sawy
مشرف / Hoda Abdel Kader Saleh
مشرف / Walaa Abdel Aty Ahmed
مشرف / Hany Mahmoud Omar Abed
مناقش / Gihan Gamal El Din Desouky
الموضوع
Oral and Maxillofacial Radiology
تاريخ النشر
2022.
عدد الصفحات
101 p. :
اللغة
الإنجليزية
الدرجة
ماجستير
التخصص
العلاج الطبيعي والرياضة والعلاج وإعادة التأهيل
تاريخ الإجازة
12/8/2022
مكان الإجازة
جامعة القاهرة - علاج طبيعي - Oral and Maxillofacial Radiology
الفهرس
Only 14 pages are availabe for public view

from 118

from 118

Abstract

To assess the accuracy of ultra-low dose and standard dose CBCT protocols in volumetric measurements of simulated bone defects in comparison with the real physical measurements
Methodology: A number of 58 defects mimicking periapical lesions were prepared on the ribs of sheep. Actual physical volumetric measurements of the defects were obtained by the water displacement method; samples were scanned by ultra-low dose and standard dose CBCT protocols. Volumetric measurements were done in each scan by two calibrated oral radiologists, with different experience, who were blinded to the parameters of the scans and to the physical measurements. The values of the CBCT volumetric measurements were compared to the physical measurements and statistically analyzed.
Results: The results of this study showed that the ultra-low dose had smaller Dahlberge error, smaller Relative Dahlberge error,tighter limits of agreement and higher concordance correlation coefficient than standard dose which means better accuracy of the segmentation procedure using the ULD protocol. A strong positive correlation was found between the two protocols and the physical measurement.
Conclusion: Ultra-low dose CBCT is suitable protocol in volumetric measurements. As both ultra-low dose protocol and standard dose protocol had a strong correlation with physical measurement, CBCT is a reliable tool in volumetric measurements.