Standardized radiometric method for evaluating phototherapy irradiance measurements and its medical applications /
الناشر
Samaa Medhat Abd Elrehem Faramawy ,
المؤلف
Samaa Medhat Abdelrehem Faramawy
هيئة الاعداد
باحث / Samaa Medhat Abdelrehem Faramawy
مشرف / Ahmed S. Monem
مشرف / Reem H. Elgebaly
مشرف / Alaa Eldin A. Abdelmageed
تاريخ النشر
2019
عدد الصفحات
200 P. :
اللغة
الإنجليزية
الدرجة
الدكتوراه
التخصص
Biophysics
تاريخ الإجازة
16/9/2019
مكان الإجازة
جامعة القاهرة - كلية العلوم - Biophysics
الفهرس
Only 14 pages are availabe for public view

from 299

from 299

Abstract

The performance of numerous phototherapy devices, developed and marketed, generally has not been adequately tested for irradiance delivery. The safety and efficiency in the operation of these devices remain one of the essential requirements for public health and concern. This thesis investigates the radiometric properties of different therapeutic devices currently used for phototherapy in Egypt. In order to achieve their effectiveness, we have established a standardized radiometric procedure based on evaluating the phototherapy devices through testing their irradiance and uniformity. About 25-50% of all full term neonates and 70-80% of prematurely newborn infants suffer from hyperbilirubinemia during their early days of life.The bilirubin molecule, which is responsible for the yellowness skin appearance of babies with jaundice, is extremely sensitive to optical energy within the spectrum radiation range 400 -500 nm.The effectiveness of phototherapy treatment depends directly on the rate of energy emitted from light source expressed in terms of irradiance W/m2. This effectiveness of neonatal hyperbilirubinaemia treatment depends directly on the wavelength and intensity (irradiance) of the light emitted during phototherapy expressed by spectral irradiance in W/m2/nm, to achieve a therapeutic reduction in the bilirubin concentration in the blood