Design and modeling of medical linear accelerators using the geant 4/GATE platform /
الناشر
Soad Mostafa Mohammed Hozayen ,
المؤلف
Soad Mostafa Mohammed Hozayen
هيئة الاعداد
باحث / Soad Mostafa Mohammed Hozayen
مشرف / Hoda Aboushady
مشرف / Hassan Fathy Ibrahim
مناقش / Badawy Abu Ibrahim Ibrahim
تاريخ النشر
2018
عدد الصفحات
160 P. :
اللغة
الإنجليزية
الدرجة
ماجستير
التخصص
Physical and Theoretical Chemistry
تاريخ الإجازة
17/3/2018
مكان الإجازة
جامعة القاهرة - كلية العلوم - Physics
الفهرس
Only 14 pages are availabe for public view

from 228

from 228

Abstract

The aim of this research is to investigate the use of GATE, the monte Carlo simulation application based on Geant 4 toolkit to estimate the percent depth doses and dose Off-Axis Ratios of a novel medical linear accelerator for photon mode field sizes 5{u00D7}5, 10{u00D7}10, 20{u00D7}20, 30{u00D7}30 and 40{u00D7}40 cm² and also for electron mode field sizes 5{u00D7}5, 10{u00D7}10, 15{u00D7}15 and 20{u00D7}20 cm² at the isocenter. A model for the 60{u00D7}50{u00D7}40.75 cm³ PTW MP3 water phantom was used in the simulations. Firstly, verification of the dose calculations of two medical linacs: Siemens ONCOR for field sizes 10{u00D7}10, 20{u00D7}20 and 30{u00D7}30 cm² and varian 600C for field sizes 3{u00D7}3, 6{u00D7}6 and 9.8{u00D7}9.8 cm² using 6 MeV electron beam, were confirmed