Surface Defects and Crystalline Structural Changes of Used Reciproc and Reciproc Blue Rotary Nickel-Titanium Files
(Vitro Study)
المؤلف
Hamza,Ahmed Mohamed .
هيئة الاعداد
باحث / أحمد محمد حمزه
مشرف / ا.د عبير عبد الحكيم الجندي
مشرف / د/ طارق مصطفي عبد العزيز
تاريخ النشر
2022
عدد الصفحات
vi(159)P:.
اللغة
الإنجليزية
الدرجة
ماجستير
التخصص
طب الأسنان
الناشر
تاريخ الإجازة
1/1/2022
مكان الإجازة
جامعة عين شمس - كلية طب الأسنان - علاج جذور
الفهرس
Only 14 pages are availabe for public view

from 159

from 159

Abstract

Summary & Conclusions
The aim of the study was to evaluate surface defects and crystalline structural changes in reciproc and reciproc blue rotary nickel-titanium files after use.
Sixty lower first molars were collected, Proximal radiographic views were taken to mesiobuccal and mesiolingual canals exhibiting a curvature, average between (25o - 35o). The method was done according to Schneider method by using Image J software.
The each tooth the mesiobuccal canal is done by reciproc file but the mesiolingual by reciproc blue file. EDTA used during the mechanical preparation. After each file the canal was irrigated by 5.25% NAOCL.
A total number of 22 files were used, 20 experimental files and 2 controlled files (one from each system for X-Ray diffraction assessment). Then each group was subdivided into 2 subgroups.
group I: 60 MB canals were prepared using 10 Reciproc files.
group II: 60 ML canals were prepared using 10 Reciproc blue files.
Classification of each group into two subgroups A & B:
Subgroup A: 10 files were used for 30 samples.
Subgroup B: the same files were used to prepare other 3 canals.
1) Scanning Electron Microscopy (SEM)
For evaluating the topography of the files, the presence or absence of the defects was checked on the surface of every file. Field Emission Scanning Electron Microscope was used with magnification from 300x to 1000x.
Results showed that there was no statistically significant difference between Reciproc and Reciproc Blue systems in new, three and six canals preparation. BUT There was a significant difference between new and after three and six canals used in disruption of cutting edge, debris and scraping.
2) X-Ray Diffraction (XRD)
Segments from the shank portions of Ni-Ti files was cut into approximately 4- 8 mm. They were adhered together and ground to obtain a plane. Then peaks was identified.
The results showed that the XRD analyses confirmed the presence of R phase in new files which decreased after use. And presence martensitc and R phase in Reciproc blue and increase R phase after use.
Conclusions:
1. Different surface defects appeared to be present in all reciproc and reciproc blue new files with various incidence rates.
2. By use for three and six times, reciproc blue files presented increased disruption of cutting edge, debris and scraping.
3. Both new and used reciproc and reciproc blue files has different phases; minor Iron phase, R-phase, and Martensite phase with various incidence rates.
4. The predominant phase in all new reciproc files is R phase and reciproc blue new files is martensite phase.
5. The predominant phase in all used reciproc is martensite and reciproc blue new files is R-phase.