Analysis of the Effect of Sampling Site of Umbilical Artery Doppler on the Variability of Umbilical Artery Pulsatility tndex /
المؤلف
Ayyad, Mohammad Samy Abd EI-Hameed.
هيئة الاعداد
باحث / Mohammad Samy Abd El-Hameed Ayyad
مشرف / Yasser Abou Talib
مشرف / Karim El-Etriby Lecturer of Obstetrics & Gyneco
مشرف / Karim El-Etriby Lecturer of Obstetrics & Gyneco
الموضوع
Laser Doppler velocimeter. Prenatal diagnosis.
تاريخ النشر
2020.
عدد الصفحات
1 VOL. (various paging’s) :
اللغة
الإنجليزية
الدرجة
ماجستير
التخصص
Multidisciplinary تعددية التخصصات
تاريخ الإجازة
1/1/2020
مكان الإجازة
جامعة حلوان - كلية الطب - Obstetrics and Gynecology
الفهرس
Only 14 pages are availabe for public view

from 118

from 118

Abstract

Objectives: The aim of the present study was to test the hypothesis that
standardising the site of sampling of umbilical artery Doppler reduces the
variability of umbilical artery Pulsatility Index (PI).
Patients and Methods: The present study was Prospective study included
120 pregnant women - with singleton fetuses of 28 week of gestation till
term and delivery-attending outpatient clinic at Helwan University
Hospitals. All pregnant females were followed monthly minimum for 3
occasions till delivery.
Results: The present study revealed the presence of significant variability
in the measurements of umbilical artery PI among the sampling sites as
well as among the different time points at which the measurements were
taken, with poor repeatability and reproducibility of the umbilical artery
PI measurements. Variagility in umbilical artery PI among the sampling
sites was demonstrated in our study. There were significant differences in
the mean umbilical artery PI among the four sampling sites, with higher
mean umbilical artery PI at PVC site, followed by AI, FL, and PL sites.
Intraclass correlation was computed to test reliability, showing poor
agreement among the four sampling sites. These findings suggest that
standardization of the site of sampling may reduce variability and increase
reliability of measurements.
Conclusion: Standardization of the site of sampling may reduce
variability and increase reliability of measurements as our study revealed
significant variability in the measurements of umbilical artery PI among
the 4 potential sampling sites as well as among the different time points at
which the measurements were taken, with poor repeatability and
reproducibility of the umbilical artery PI measurements.
Keywords: Sampling Site, Umbilical Artery Doppler, Variability,
Umbilical Artery Pulsatility Index
.